Andatech Precision+ user manual

Thank you for purchasing an Andatech Precision+ breathalyser

The operator of this device should read the operation manual and be familiar with the device operation before using it. Retain the manual for future reference.

View and download the Andatech Precision+ breathalyser user manual [PDF].


Components

Andatech Precision+ breathalyser components

  1. Mouthpiece holder
  2. LCD Screen
  3. On/Off button
  4. Function button
  5. USB Port (Not used)
  6. Battery compartment cover

Preparation

Gently pull down the battery compartment cover to remove it. Insert fresh alkaline batteries (4 x AAA), checking that the + and - on each battery are aligned according to the polarity diagram inside. Carefully replace the battery compartment cover.

Switching the unit on and off

Press and hold the ON/OFF button for 2-3 seconds to switch the unit on or off.
The unit will automatically switch off after 2 minutes of non-use.

Number of tests

Andatech Precision+ breathalyser number of tests

When the unit first starts up, it will display the number of tests taken and the number of tests remaining before calibration is required.

Normal operation

Andatech Precision+ breathalyser normal operation

  1. Insert a disposable mouthpiece through the unit.
  2. Switch on the unit. When the screen reads BLOW, take a deep breath and blow steadily into the mouthpiece for at least 4 seconds.
  3. Continue blowing until the beep stops. You will hear a click.
  4. Once complete, the unit displays the reading on the screen.
  •  The next test can be performed by pressing the ON/OFF button once, then repeating the previous steps. If the reading from the previous test was higher than zero, please wait 3 minutes before performing the next test to ensure accurate results.

Passive test

Andatech Precision+ breathalyser passive test

  1. No mouthpieces are required for this operation.
  2. When the screen reads BLOW, take a deep breath and blow gently into the hole for 3 seconds. While blowing press the ON/OFF button to record a breath sample. Once complete, you will hear a click.
  3. Once complete, the unit displays the reading on the screen as No Alcohol or Alcohol Detected.
  4. To conduct another test straight away, press the ON/OFF button once and repeat previous steps.
  5. If a more specific reading is required, use a mouthpiece and test under normal operation.

Troubleshooting

Error "Please blow harder"

Breath sample was not long, hard, or steady enough, or you did not wait for the beep to end before you stopped blowing. Blow a long, slow and steady breath. Don't stop blowing until the beep ends.

'Calibration Needed' warning

The unit has passed its calibration due date or has exceeded the number of allowed tests. You can press any button to continue using the unit, however we strongly recommend that you send it back to us for calibration.

Battery life too short

Use only alkaline batteries, or if using rechargeable batteries, use either NiMH or rechargeable alkaline batteries.

Measurements & Conditions for Test

  • Use a clean mouthpiece for each test.
  • Do not allow samples containing smoke, tobacco or other chemical products to enter the unit, as damage to the fuel cell can occur.
  • There must be an interval of at least 15 minutes after alcohol consumption as alcohol residue in the mouth may give a false reading.
    Such residue may also be left by aromatic drinks, mouthwash, medicines and drops, as well as by belching and vomiting. Rinsing your mouth with water or non-alcoholic drinks does not substitute for an interval.
  • A person’s blood alcohol concentration (BAC) can continue to rise for up to 2 hours after drinking alcohol. If a caution results or a result is close to the designated BAC, it is recommended to wait and retest after 20 minutes. It can take 10 hours or more for a person’s BAC to return to zero after a high BAC has been reached.
  • A further test should be carried out later in the day or the following morning.
  • Working temperature range: 5°C to 40°C.
    Storage temperature: -15°C to 65°C.

Liability for proper function or damage

The liability for proper functioning of the instrument is irrevocably transferred to the owner or operator in the event the instrument is serviced or repaired by personnel not employed or authorised by Andatech. In such an event, the warranty will be void.

Disclaimers

  • This product must be used strictly in accordance with the instructions in this manual.
  • This product is intended to provide the user with an indication of alcohol on the breath. However, neither the manufacturer nor the supplier gives any warranty (express or implied) that the product can give any indication of blood alcohol level.
  • A person’s BAC (Blood Alcohol Concentration) can only be accurately measured by the administration of a blood test. This product only measures breath alcohol content which, in turn, only gives an approximate indication of blood alcohol concentration.
  • In these circumstances the user of this product should not rely on the information provided thereby as an accurate indication of his blood alcohol concentration or of his level of intoxication or of his fitness to drive and in particular should not take any action in reliance on such information as the product can provide no more than an indication of alcohol on the breath.
  • If the user of this product makes any decision or takes any action based on the use of the product such decision or such action shall be at the sole risk of the user and neither the manufacturer nor the supplier assumes any responsibility for such decision or action. In particular, neither the manufacturer nor the supplier shall be in any way liable if the user of the product asserts that the product led him to believe that he was fit to operate a motor vehicle and is subsequently found not to be so fit.
  • In particular, neither the manufacturer nor the supplier shall be in any way liable if the user of the product asserts that the product led them to believe they were fit to operate a motor vehicle and are subsequently found not to be so fit.

Things to remember

  1. Always wait a minimum of 15 minutes after your last drink before testing.
  2. NEVER take a sip of alcohol and blow immediately into the device.
  3. Ignoring these two points WILL cause damage to the unit that is NOT covered by warranty.

Recalibration and service

Andatech recommends the Andatech® Precision Plus be calibrated every 1,000 tests or 6 months (whichever comes first).

For recalibration and technical support, please go to my.andatech.com.au or contact Andatech support.